Study Objectives: <p>The Belun ring is a new home sleep apnea testing device using a pulse oximeter sensor and a neural network algorithm, but its data in children are limited. This study aims to evaluate the correlation and agreement of the Belun ring, compared with polysomnography (PSG) and determine the diagnostic accuracy of the Belun ring for moderate-to-severe obstructive sleep apnea (OSA).</p> Methods: <p>This is a cross-sectional observational study in children aged 5–18 years with suspected OSA between June 2023 and February 2024. The Belun ring and PSG were undertaken on eligible participants to assess apnea-hypopnea index (AHI) in the same sleep test session.</p> Results: <p>Of 75 children enrolled, OSA was diagnosed in 74 children by PSG. The Belun AHI (B-AHI) was moderately correlated with the PSG AHI (P-AHI) (<i>r =</i> .63, <i>P</i> &lt; .001) with mean difference (standard deviation) −7.8 (13.91) events/h. The area under the receiver operating characteristic curve of the B-AHI to identify moderate-to-severe OSA (P-AHI &gt; 5 events/h) was 0.66, and the B-AHI cut-off of 3 events/h yielded 74.1% sensitivity and 52.4% specificity. The B-AHI cut-off of 2 events/h yielded 92.6% sensitivity, and 7 events/h yielded 95.2% specificity.</p> Conclusions: <p>Despite the correlation, the difference in AHI between the Belun ring and PSG in children was noted. Either single or multiple B-AHI cut-offs to diagnose, include or exclude moderate-to-severe OSA may be valuable, but their implementation must be approached with caution.</p> Clinical Trial Registration: <p>Registry: Thai Clinical Trials Registry; Name: Diagnostic Accuracy of the Belun Ring in Children at Risk of Obstructive Sleep Apnea; URL: <a href="https://www.thaiclinicaltrials.org/show/TCTR20240604003">https://www.thaiclinicaltrials.org/show/TCTR20240604003</a>; Identifier: TCTR20240604003.</p> Citation: <p>Panichapat N, Niyomkarn W, Boonjindasup W, Thiamrakij P, Sritippayawan S, Deerojanawong J. Diagnostic accuracy of the Belun ring in children at risk of obstructive sleep apnea. <i>J Clin Sleep Med</i>. 2025;21(1):123–128.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Diagnostic accuracy of the Belun ring in children at risk of obstructive sleep apnea

  • Nuttida Panichapat,
  • Watit Niyomkarn,
  • Wicharn Boonjindasup,
  • Pariyapa Thiamrakij,
  • Suchada Sritippayawan,
  • Jitladda Deerojanawong

摘要

Study Objectives:

The Belun ring is a new home sleep apnea testing device using a pulse oximeter sensor and a neural network algorithm, but its data in children are limited. This study aims to evaluate the correlation and agreement of the Belun ring, compared with polysomnography (PSG) and determine the diagnostic accuracy of the Belun ring for moderate-to-severe obstructive sleep apnea (OSA).

Methods:

This is a cross-sectional observational study in children aged 5–18 years with suspected OSA between June 2023 and February 2024. The Belun ring and PSG were undertaken on eligible participants to assess apnea-hypopnea index (AHI) in the same sleep test session.

Results:

Of 75 children enrolled, OSA was diagnosed in 74 children by PSG. The Belun AHI (B-AHI) was moderately correlated with the PSG AHI (P-AHI) (r = .63, P < .001) with mean difference (standard deviation) −7.8 (13.91) events/h. The area under the receiver operating characteristic curve of the B-AHI to identify moderate-to-severe OSA (P-AHI > 5 events/h) was 0.66, and the B-AHI cut-off of 3 events/h yielded 74.1% sensitivity and 52.4% specificity. The B-AHI cut-off of 2 events/h yielded 92.6% sensitivity, and 7 events/h yielded 95.2% specificity.

Conclusions:

Despite the correlation, the difference in AHI between the Belun ring and PSG in children was noted. Either single or multiple B-AHI cut-offs to diagnose, include or exclude moderate-to-severe OSA may be valuable, but their implementation must be approached with caution.

Clinical Trial Registration:

Registry: Thai Clinical Trials Registry; Name: Diagnostic Accuracy of the Belun Ring in Children at Risk of Obstructive Sleep Apnea; URL: https://www.thaiclinicaltrials.org/show/TCTR20240604003; Identifier: TCTR20240604003.

Citation:

Panichapat N, Niyomkarn W, Boonjindasup W, Thiamrakij P, Sritippayawan S, Deerojanawong J. Diagnostic accuracy of the Belun ring in children at risk of obstructive sleep apnea. J Clin Sleep Med. 2025;21(1):123–128.