Abstract <p>We discuss a possibility of using a xenon gamma-ray spectrometer (XGRS) based on a&#xa0;cylindrical ionization chamber filled with xenon at a pressure of 40 atm for measuring dose loads during neutron capture therapy (NCT). Advantages of the detector include its resistance to neutron fluxes and a high energy resolution (1.7% at gamma-quanta energies of 662 keV). These characteristics provide the capability to obtain reliable and effective results during NCT. Use is also made of a Pu‒Be neutron source and a pulsed neutron generator.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Determination of the Dose Load in Neutron Capture Therapy Using a Xenon Gamma-Ray Spectrometer

  • I. F. Khimmatov,
  • S. E. Ulin

摘要

Abstract

We discuss a possibility of using a xenon gamma-ray spectrometer (XGRS) based on a cylindrical ionization chamber filled with xenon at a pressure of 40 atm for measuring dose loads during neutron capture therapy (NCT). Advantages of the detector include its resistance to neutron fluxes and a high energy resolution (1.7% at gamma-quanta energies of 662 keV). These characteristics provide the capability to obtain reliable and effective results during NCT. Use is also made of a Pu‒Be neutron source and a pulsed neutron generator.