Background <p>After its first ex-vivo and in-vivo demonstration, Positronium Lifetime Imaging (PLI) has received considerable interest as a potential new diagnostic biomarker. High sensitivity Positron Emission Tomography (PET) systems are needed for PLI since it requires simultaneous registration of annihilation photons and prompt gamma. In this simulation-based study, a&#xa0;feasibility of PLI with the long axial field-of-view Biograph Vision Quadra (Quadra) and the Total Body J-PET scanner was investigated.</p> Methods <p>The study was performed using the GATE software. Background radiation, present within the Quadra tomograph, was added to the simulation. First, the optimal placement of the energy window for the registration of the prompt gamma was investigated. Next, the organ-wise sensitivity of Quadra was calculated for the <InlineEquation ID="IEq1"><EquationSource Format="TEX">\(^{68}\)</EquationSource></InlineEquation>Ga, <InlineEquation ID="IEq2"><EquationSource Format="TEX">\(^{44}\)</EquationSource></InlineEquation>Sc, <InlineEquation ID="IEq3"><EquationSource Format="TEX">\(^{22}\)</EquationSource></InlineEquation>Na and <InlineEquation ID="IEq4"><EquationSource Format="TEX">\(^{124}\)</EquationSource></InlineEquation>I radioisotopes. Finally, the sensitivity for the scandium isotope was compared to the sensitivities obtainable with the Total Body J-PET scanner, as well as with the modular J-PET prototype.</p> Results <p>The PLI sensitivities for the Quadra with the background radiation are estimated to 9.22(3), 10.46(4), 5.91(3), and 15.39(4) cps/kBq for the <InlineEquation ID="IEq5"><EquationSource Format="TEX">\(^{44}\)</EquationSource></InlineEquation>Sc, <InlineEquation ID="IEq6"><EquationSource Format="TEX">\(^{68}\)</EquationSource></InlineEquation>Ga, <InlineEquation ID="IEq7"><EquationSource Format="TEX">\(^{22}\)</EquationSource></InlineEquation>Na and <InlineEquation ID="IEq8"><EquationSource Format="TEX">\(^{124}\)</EquationSource></InlineEquation>I radioisotopes, respectively. The highest sensitivity was obtained when the energy window for the deexcitation photon is adjacent to the energy window for the annihilation photons. In case of J-PET, the sensitivities for the <InlineEquation ID="IEq9"><EquationSource Format="TEX">\(^{44}\)</EquationSource></InlineEquation>Sc were estimated to 0.062(08) and 1.714(40) cps/kBq for the modular and Total Body J-PET scanners, respectively. The determined PLI sensitivities with Quadra and the Total Body J-PET are in the order of sensitivities of metabolic PET imaging with the short axial field-of-view (<InlineEquation ID="IEq10"><EquationSource Format="TEX">\(\sim \)</EquationSource></InlineEquation>20&#xa0;cm) PET scanners based on pure positron emitter isotopes.</p> Conclusions <p>The PLI sensitivity of Quadra has been computed for the <InlineEquation ID="IEq11"><EquationSource Format="TEX">\(^{68}\)</EquationSource></InlineEquation>Ga, <InlineEquation ID="IEq12"><EquationSource Format="TEX">\(^{44}\)</EquationSource></InlineEquation>Sc, <InlineEquation ID="IEq13"><EquationSource Format="TEX">\(^{22}\)</EquationSource></InlineEquation>Na and <InlineEquation ID="IEq14"><EquationSource Format="TEX">\(^{124}\)</EquationSource></InlineEquation>I radioisotopes. A sensitivity gain by a factor of 150 was estimated relative to the modular J-PET system previously used for the first in-vivo PLI. Additionally, the sensitivity for Total Body J-PET is expected to be about 30 times higher than for modular J-PET.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Feasibility simulation study of positronium lifetime imaging with the Biograph Vision Quadra and J-PET tomographs

  • Szymon Parzych,
  • Szymon Niedźwiecki,
  • Ermias Yitayew Beyene,
  • Neha Chug,
  • Maurizio Conti,
  • Catalina Curceanu,
  • Eryk Czerwiński,
  • Manish Das,
  • Kavya Valsan Eliyan,
  • Jakub Hajduga,
  • Sharareh Jalali,
  • Krzysztof Kacprzak,
  • Tevfik Kaplanoglu,
  • Łukasz Kapłon,
  • Kamila Kasperska,
  • Aleksander Khreptak,
  • Grzegorz Korcyl,
  • Tomasz Kozik,
  • Deepak Kumar,
  • Anoop Kunimmal Venadan,
  • Karol Kubat,
  • Edward Lisowski,
  • Filip Lisowski,
  • Justyna Medrala-Sowa,
  • Wiktor Mryka,
  • Simbarashe Moyo,
  • Piyush Pandey,
  • Elena Perez del Rio,
  • Alessio Porcelli,
  • Bartłomiej Rachwał,
  • Martin Rädler,
  • Axel Rominger,
  • Sushil Sharma,
  • Kuangyu Shi,
  • Magdalena Skurzok,
  • William M. Steinberger,
  • Tomasz Szumlak,
  • Pooja Tanty,
  • Keyvan Tayefi Ardebili,
  • Satyam Tiwari,
  • Ewa Ł. Stepień,
  • Paweł Moskal

摘要

Background

After its first ex-vivo and in-vivo demonstration, Positronium Lifetime Imaging (PLI) has received considerable interest as a potential new diagnostic biomarker. High sensitivity Positron Emission Tomography (PET) systems are needed for PLI since it requires simultaneous registration of annihilation photons and prompt gamma. In this simulation-based study, a feasibility of PLI with the long axial field-of-view Biograph Vision Quadra (Quadra) and the Total Body J-PET scanner was investigated.

Methods

The study was performed using the GATE software. Background radiation, present within the Quadra tomograph, was added to the simulation. First, the optimal placement of the energy window for the registration of the prompt gamma was investigated. Next, the organ-wise sensitivity of Quadra was calculated for the \(^{68}\)Ga, \(^{44}\)Sc, \(^{22}\)Na and \(^{124}\)I radioisotopes. Finally, the sensitivity for the scandium isotope was compared to the sensitivities obtainable with the Total Body J-PET scanner, as well as with the modular J-PET prototype.

Results

The PLI sensitivities for the Quadra with the background radiation are estimated to 9.22(3), 10.46(4), 5.91(3), and 15.39(4) cps/kBq for the \(^{44}\)Sc, \(^{68}\)Ga, \(^{22}\)Na and \(^{124}\)I radioisotopes, respectively. The highest sensitivity was obtained when the energy window for the deexcitation photon is adjacent to the energy window for the annihilation photons. In case of J-PET, the sensitivities for the \(^{44}\)Sc were estimated to 0.062(08) and 1.714(40) cps/kBq for the modular and Total Body J-PET scanners, respectively. The determined PLI sensitivities with Quadra and the Total Body J-PET are in the order of sensitivities of metabolic PET imaging with the short axial field-of-view (\(\sim \)20 cm) PET scanners based on pure positron emitter isotopes.

Conclusions

The PLI sensitivity of Quadra has been computed for the \(^{68}\)Ga, \(^{44}\)Sc, \(^{22}\)Na and \(^{124}\)I radioisotopes. A sensitivity gain by a factor of 150 was estimated relative to the modular J-PET system previously used for the first in-vivo PLI. Additionally, the sensitivity for Total Body J-PET is expected to be about 30 times higher than for modular J-PET.