Feasibility simulation study of positronium lifetime imaging with the Biograph Vision Quadra and J-PET tomographs
摘要
After its first ex-vivo and in-vivo demonstration, Positronium Lifetime Imaging (PLI) has received considerable interest as a potential new diagnostic biomarker. High sensitivity Positron Emission Tomography (PET) systems are needed for PLI since it requires simultaneous registration of annihilation photons and prompt gamma. In this simulation-based study, a feasibility of PLI with the long axial field-of-view Biograph Vision Quadra (Quadra) and the Total Body J-PET scanner was investigated.
MethodsThe study was performed using the GATE software. Background radiation, present within the Quadra tomograph, was added to the simulation. First, the optimal placement of the energy window for the registration of the prompt gamma was investigated. Next, the organ-wise sensitivity of Quadra was calculated for the
The PLI sensitivities for the Quadra with the background radiation are estimated to 9.22(3), 10.46(4), 5.91(3), and 15.39(4) cps/kBq for the
The PLI sensitivity of Quadra has been computed for the