Background <p>Clear aligners face biomechanical limitations in complex tooth movements, particularly first molar mesialization, despite attachment use. Finite element analysis (FEA) can elucidate optimal attachment designs for force delivery.</p> Objectives <p>To compare displacement patterns and stress distribution during maxillary first molar mesialization using four attachment designs via 3D FEA.</p> Methods <p>A maxillary model (MRI-derived) was created with periodontal ligament (PDL), alveolar bone, and clear aligner (0.75&#xa0;mm thickness). Five scenarios were simulated: no attachment (Model-I), vertical rectangular (Model-II), horizontal rectangular (Model-III), optimized double semi-ellipsoidal (Model-IV), and Yin-Yang attachments (Model-V). Mesial displacement (0.5&#xa0;mm) was applied, and deformation/stress were analyzed using ANSYS Workbench.</p> Results <p>Model-IV (optimized attachment) demonstrated the least first molar tipping, evidenced by its lowest total mesial crown displacement (0.319&#xa0;mm), and provided the best mesiodistal control. Conversely, Model-I (no attachment) exhibited the highest tipping, with a total mesial crown displacement of 0.376&#xa0;mm. Yin-Yang attachments (Model-V) significantly reduced buccal displacement by 92% compared to Model-I. Horizontal rectangular attachments minimized rotational movement.</p> Conclusions <p>Optimized double semi-ellipsoidal attachments provide superior first molar mesialization control, while Yin-Yang designs enhance buccal-lingual stability. Attachment geometry critically influences aligner efficacy.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Biomechanical effects of attachment designs in maxillary first molar mesialization with clear aligners: a finite element study

  • Resul Vatansever,
  • Hakan Gurcan Gurel,
  • Osman Koc

摘要

Background

Clear aligners face biomechanical limitations in complex tooth movements, particularly first molar mesialization, despite attachment use. Finite element analysis (FEA) can elucidate optimal attachment designs for force delivery.

Objectives

To compare displacement patterns and stress distribution during maxillary first molar mesialization using four attachment designs via 3D FEA.

Methods

A maxillary model (MRI-derived) was created with periodontal ligament (PDL), alveolar bone, and clear aligner (0.75 mm thickness). Five scenarios were simulated: no attachment (Model-I), vertical rectangular (Model-II), horizontal rectangular (Model-III), optimized double semi-ellipsoidal (Model-IV), and Yin-Yang attachments (Model-V). Mesial displacement (0.5 mm) was applied, and deformation/stress were analyzed using ANSYS Workbench.

Results

Model-IV (optimized attachment) demonstrated the least first molar tipping, evidenced by its lowest total mesial crown displacement (0.319 mm), and provided the best mesiodistal control. Conversely, Model-I (no attachment) exhibited the highest tipping, with a total mesial crown displacement of 0.376 mm. Yin-Yang attachments (Model-V) significantly reduced buccal displacement by 92% compared to Model-I. Horizontal rectangular attachments minimized rotational movement.

Conclusions

Optimized double semi-ellipsoidal attachments provide superior first molar mesialization control, while Yin-Yang designs enhance buccal-lingual stability. Attachment geometry critically influences aligner efficacy.