Background <p>The diagnostic yield of conventional gastrointestinal (GI) endoscopy for early cancers is low because it is mainly based on morphological changes of tumors. Molecular functional changes in tumors precede morphological changes. Cherenkov luminescence endoscopy (CLE) system can perform molecular imaging of GI cancers, achieving early diagnosis of cancers. However, previous CLE systems had only been able to detect Cherenkov luminescence (CL) from about one μCi nuclide at a minimum (in vivo), but the nuclide probe absorbed by the tumor of a patient was often much less than one μCi at a routinely administered dose. This study aims to construct a clinically usable high-sensitivity CLE for molecular imaging of GI cancers.</p> Results <p>The minimum resolvable radioactivity of the CLE reached 0.020&#xa0;μCi within 300&#xa0;s (in vivo), with a sensitivity at the nanocurie for the first time. The detection sensitivity of the CLE increased by up to nearly twenty-two times over the previous system. In tumor-bearing nude mice, CLE could effectively identify all tumors with 100% concordance with both histopathology and PET/CT, and the CL signals of tumors were much stronger than those of the surrounding normal tissues (<i>P</i> &lt; 0.05). The quality of CLE imaging at 60&#xa0;s was comparable to that at 300&#xa0;s (signal-to-background ratio, 2.70 ± 0.48 versus 2.98 ± 0.69, <i>P</i> = 0.56).</p> Conclusions <p>We constructed a high-sensitivity CLE that could detect radionuclides at the nanocurie radioactivity. The CLE could detect cancers accurately through rapid molecular imaging and had the potential for early diagnosis of GI cancers in clinical practice.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Construction of a high-sensitivity Cherenkov luminescence endoscopy system for the detection of gastrointestinal cancers

  • Ze Yang,
  • Tian-Tian Pang,
  • Zhuo-Jun Wu,
  • Tian-Yu Yan,
  • Jing-Min Yu,
  • Xin-Yu Wang,
  • Dan Liu,
  • Xiao-Jian Lu,
  • Xiao-Yu Kang,
  • Gui-Yu Li,
  • Cheng Bai,
  • Xiao-Juan Xi,
  • Zu-Hong Tian,
  • Yu Qi,
  • Ming-Ru Zhang,
  • Fei Kang,
  • Jing Wang,
  • Xue-Li Chen,
  • Kai-Chun Wu

摘要

Background

The diagnostic yield of conventional gastrointestinal (GI) endoscopy for early cancers is low because it is mainly based on morphological changes of tumors. Molecular functional changes in tumors precede morphological changes. Cherenkov luminescence endoscopy (CLE) system can perform molecular imaging of GI cancers, achieving early diagnosis of cancers. However, previous CLE systems had only been able to detect Cherenkov luminescence (CL) from about one μCi nuclide at a minimum (in vivo), but the nuclide probe absorbed by the tumor of a patient was often much less than one μCi at a routinely administered dose. This study aims to construct a clinically usable high-sensitivity CLE for molecular imaging of GI cancers.

Results

The minimum resolvable radioactivity of the CLE reached 0.020 μCi within 300 s (in vivo), with a sensitivity at the nanocurie for the first time. The detection sensitivity of the CLE increased by up to nearly twenty-two times over the previous system. In tumor-bearing nude mice, CLE could effectively identify all tumors with 100% concordance with both histopathology and PET/CT, and the CL signals of tumors were much stronger than those of the surrounding normal tissues (P < 0.05). The quality of CLE imaging at 60 s was comparable to that at 300 s (signal-to-background ratio, 2.70 ± 0.48 versus 2.98 ± 0.69, P = 0.56).

Conclusions

We constructed a high-sensitivity CLE that could detect radionuclides at the nanocurie radioactivity. The CLE could detect cancers accurately through rapid molecular imaging and had the potential for early diagnosis of GI cancers in clinical practice.