Background <p>Accurate identification and treatment of hypoglycemia or low blood glucose is important for neonates, especially premature, small for gestational age, and intra-uterine growth restricted (“small and sick”) newborns. Existing point-of-care (POC) adult glucometers, intended for home management of adults or children with diabetes, are widely available and less costly than machines designed and validated for neonates, but they must be validated at the lower glucose levels relevant for neonates and at appropriate hematocrit levels before being used for neonates.</p> Methods <p>We simulated neonatal blood by altering hematocrit (to 18%, 40%, and 55%) and glucose concentration (to 40&#xa0;mg/dL and 150&#xa0;mg/dL) of normal adult blood. We then used 11 POC glucometers to measure glucose levels in these samples and compared results to values measured with a clinical chemistry analyzer (YSI 2300).</p> Results <p>We report mean percent bias, mean absolute relative difference, and standard deviation for each tested condition and glucometer. We ranked the tested glucometers based on a combination of accuracy and precision for measuring neonatal glucose levels and consumable cost. From best-performing to worst, they are Group 1: StatStrip and StatStrip Xpress 2; Group 2: AccuChek Guide, AccuChek GuideMe, AccuChek Instant, and AccuChek Performa; Group 3: AccuChek Active, GlucoNavii; Group 4: HemoCue RT, Nipro Premier, and OneTouchUltra 2.</p> Conclusions <p>This paper describes a simple, laboratory-based method to test glucometers across a wide range of hematocrit values and glucose concentrations and contributes up-to-date testing data on currently available POC glucometers at ranges relevant to neonatal use.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Evaluation of commercial point-of-care glucometers for detection and monitoring of neonatal hypoglycemia in resource-constrained settings

  • Meaghan Bond,
  • Elizabeth Asma,
  • Joseph Peterson,
  • Lauren Oparah,
  • Millicent Alooh,
  • Danica Kumara,
  • Z. Maria Oden,
  • Chinyere Ezeaka,
  • Elizabeth Molyneux,
  • Rebecca Richards-Kortum

摘要

Background

Accurate identification and treatment of hypoglycemia or low blood glucose is important for neonates, especially premature, small for gestational age, and intra-uterine growth restricted (“small and sick”) newborns. Existing point-of-care (POC) adult glucometers, intended for home management of adults or children with diabetes, are widely available and less costly than machines designed and validated for neonates, but they must be validated at the lower glucose levels relevant for neonates and at appropriate hematocrit levels before being used for neonates.

Methods

We simulated neonatal blood by altering hematocrit (to 18%, 40%, and 55%) and glucose concentration (to 40 mg/dL and 150 mg/dL) of normal adult blood. We then used 11 POC glucometers to measure glucose levels in these samples and compared results to values measured with a clinical chemistry analyzer (YSI 2300).

Results

We report mean percent bias, mean absolute relative difference, and standard deviation for each tested condition and glucometer. We ranked the tested glucometers based on a combination of accuracy and precision for measuring neonatal glucose levels and consumable cost. From best-performing to worst, they are Group 1: StatStrip and StatStrip Xpress 2; Group 2: AccuChek Guide, AccuChek GuideMe, AccuChek Instant, and AccuChek Performa; Group 3: AccuChek Active, GlucoNavii; Group 4: HemoCue RT, Nipro Premier, and OneTouchUltra 2.

Conclusions

This paper describes a simple, laboratory-based method to test glucometers across a wide range of hematocrit values and glucose concentrations and contributes up-to-date testing data on currently available POC glucometers at ranges relevant to neonatal use.