Background <p>Recent HD systems have moved from pre-defined dialysate flow rates (e.g. 300, 500, 800 mL/min) to an individual setting, or an automatic adjustment with a fixed factor of the dialysate flow to the blood flow. Both allows more flexibility of choosing the appropriate dialysate flow as an optimal adjustment to the current blood flow.</p> Aim of the work <p>To study the possible effects of using autoflow factor function on hemodialysis (HD) efficacy in regular hemodialysis patients.</p> Methods <p>A prospective cross section study included 55 ESRD patients on hemodialysis and conducted in hemodialysis unit in Menoufia university hospital.</p> Results <p>There was no statistically significant difference between the dialysate flow rate 500 mL/min and auto Flow factor regarding the Kt/V (<i>p</i> = 0.068).</p> Conclusion <p>The present study demonstrates that the use of the Auto Flow Factor Function does not result in a statistically significant difference in dialysis efficacy compared to a fixed dialysate flow rate of 500 mL/min. Given the potential for dialysate and energy savings, the implementation of Auto Flow may represent a step toward environmentally sustainable dialysis without compromising patient care.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Effects of using auto flow factor function on hemodialysis efficacy in regular hemodialysis patients

  • Said S. A. Khamis,
  • Ahmed Ragheeb Tawfeek,
  • Mohammed Fathy Ragab,
  • Heba Kamal Abd elkhalk

摘要

Background

Recent HD systems have moved from pre-defined dialysate flow rates (e.g. 300, 500, 800 mL/min) to an individual setting, or an automatic adjustment with a fixed factor of the dialysate flow to the blood flow. Both allows more flexibility of choosing the appropriate dialysate flow as an optimal adjustment to the current blood flow.

Aim of the work

To study the possible effects of using autoflow factor function on hemodialysis (HD) efficacy in regular hemodialysis patients.

Methods

A prospective cross section study included 55 ESRD patients on hemodialysis and conducted in hemodialysis unit in Menoufia university hospital.

Results

There was no statistically significant difference between the dialysate flow rate 500 mL/min and auto Flow factor regarding the Kt/V (p = 0.068).

Conclusion

The present study demonstrates that the use of the Auto Flow Factor Function does not result in a statistically significant difference in dialysis efficacy compared to a fixed dialysate flow rate of 500 mL/min. Given the potential for dialysate and energy savings, the implementation of Auto Flow may represent a step toward environmentally sustainable dialysis without compromising patient care.