Objective <p>To evaluate the efficacy of single-energy metal artifact reduction (SEMAR) algorithm in removing metallic artifacts for patients with novel leadless pacemakers (PMs) compared to those with conventional transvenous PMs.</p> Methods <p>30 patients with transvenous PMs and 30 patients with leadless PMs having coronary CT angiography (CCTA) were included. Images were reconstructed using two methods: a conventional hybrid-iterative reconstruction (HIR) algorithm separately, and the HIR combined with SEMAR algorithm. The artifact index (AI) was calculated for both datasets to quantitatively assess the artifact reduction. Two senior radiologists independently scored the artifact extent and diagnostic confidence of surrounding tissues. The position of artifacts-impaired tissues was also visually evaluated as follows: non-impaired, impaired, and equivocal. The study was ethics committee-approved.</p> Results <p>The leadless PM<sub>SEMAR</sub> exhibited significantly lower AIs than the leadless PM<sub>non−SEMAR</sub>, with an improvement of hypodense artifacts and a significant reduction of hyperdense artifacts (all <i>p</i> &lt; 0.001). The AIs of leadless PMs performed with and without SEMAR were slightly lower than those of transvenous PMs, although no statistical difference existed (all <i>p</i> &gt; 0.05). Qualitatively, the artifact extent of SEMAR significantly outperformed non-SEMAR in both types of PM (<i>p</i> &lt; 0.001). The aorta, right coronary artery, right atrium, and left ventricle were less impaired in leadless PM<sub>non−SEMAR</sub> compared to transvenous PM<sub>non−SEMAR</sub> (all <i>p</i> &lt; 0.05).</p> Conclusions <p>SEMAR provides effective reduction of metallic artifacts from both types of cardiac PMs. The artifact-impaired tissues are fewer on leadless PMs, resulting in an enhanced diagnostic confidence in the surrounding tissues.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Reduction of metallic artifacts from the novel leadless pacemaker using single-energy metal artifact reduction algorithm in coronary CT angiography

  • Yahui Hu,
  • Ying Chen,
  • Jianjun Huo,
  • Linxin Zhang,
  • Yan Li,
  • Qian Liu,
  • Yaohui Yu,
  • Haicheng Qi,
  • Zicheng Zhao,
  • Jiaojiao Zhang,
  • Min Xu,
  • Yan Xing

摘要

Objective

To evaluate the efficacy of single-energy metal artifact reduction (SEMAR) algorithm in removing metallic artifacts for patients with novel leadless pacemakers (PMs) compared to those with conventional transvenous PMs.

Methods

30 patients with transvenous PMs and 30 patients with leadless PMs having coronary CT angiography (CCTA) were included. Images were reconstructed using two methods: a conventional hybrid-iterative reconstruction (HIR) algorithm separately, and the HIR combined with SEMAR algorithm. The artifact index (AI) was calculated for both datasets to quantitatively assess the artifact reduction. Two senior radiologists independently scored the artifact extent and diagnostic confidence of surrounding tissues. The position of artifacts-impaired tissues was also visually evaluated as follows: non-impaired, impaired, and equivocal. The study was ethics committee-approved.

Results

The leadless PMSEMAR exhibited significantly lower AIs than the leadless PMnon−SEMAR, with an improvement of hypodense artifacts and a significant reduction of hyperdense artifacts (all p < 0.001). The AIs of leadless PMs performed with and without SEMAR were slightly lower than those of transvenous PMs, although no statistical difference existed (all p > 0.05). Qualitatively, the artifact extent of SEMAR significantly outperformed non-SEMAR in both types of PM (p < 0.001). The aorta, right coronary artery, right atrium, and left ventricle were less impaired in leadless PMnon−SEMAR compared to transvenous PMnon−SEMAR (all p < 0.05).

Conclusions

SEMAR provides effective reduction of metallic artifacts from both types of cardiac PMs. The artifact-impaired tissues are fewer on leadless PMs, resulting in an enhanced diagnostic confidence in the surrounding tissues.