Abstract <p>The requirements for sensors of low- and medium-field magnetic resonance imaging (MRI) systems are analyzed. The design of the transmitting coil for a 0.4 T magnetic-field induction MRI system is described. Special elements improving the uniformity of the generated RF field have been included in the coil. The dimensions of the coil elements have been numerically simulated and optimized. The RF field distribution inside the coil has been calculated and experimentally measured using a coil prototype having the optimal dimensions. Experiments have been carried out on the MRI system to evaluate the RF field uniformity.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

The Receiving and Transmitting System with an Improved Field Uniformity for the Magnetic Resonance Imaging System with a Magnetic Field Induction of 0.4 T

  • V. L. Odivanov,
  • Ya. V. Fattakhov,
  • A. A. Bayazitov,
  • A. R. Fakhrutdinov,
  • V. A. Shagalov

摘要

Abstract

The requirements for sensors of low- and medium-field magnetic resonance imaging (MRI) systems are analyzed. The design of the transmitting coil for a 0.4 T magnetic-field induction MRI system is described. Special elements improving the uniformity of the generated RF field have been included in the coil. The dimensions of the coil elements have been numerically simulated and optimized. The RF field distribution inside the coil has been calculated and experimentally measured using a coil prototype having the optimal dimensions. Experiments have been carried out on the MRI system to evaluate the RF field uniformity.