<p>Portable low-field MRI offers an affordable and mobile alternative to conventional high-field scanners, enabling imaging in point-of-care and resource-limited settings. However, its lower signal-to-noise ratio, reduced resolution, and acquisition artifacts raise concerns about the accuracy of standard image registration methods. Reliable nonlinear registration is critical for a wide range of emerging applications, including frequent brain monitoring, assessment of neurodegenerative disease progression, and evaluation of treatment effects such as those associated with Alzheimer’s therapeutics. In this work, we systematically evaluated state-of-the-art registration approaches on simulated low-field scans (obtained by downsampling high-field images) and on real low-field brain MRI data. We compared three approaches that are representative of current registration methods: classical optimization (NiftyReg), learning-based registration (SynthMorph), and synthesis-based registration (SynthSR+NiftyReg). Using downsampled high-field scans, all methods performed well, achieving high Dice scores and smooth deformation fields, indicating that reduced resolution alone does not hinder registration. In contrast, real low-field data exhibited lower accuracy, primarily due to geometric distortion and other acquisition-specific artifacts. Among the tested approaches, the synthesis-based pipeline achieved the most robust performance across subjects and modalities. Overall, existing algorithms can accommodate resolution limitations, however, future methods could further enhance coregistration by explicitly addressing the distortions present in low-field MRI scans.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

On the accuracy of nonlinear image registration in portable low-field 3D brain MRI

  • J. Eugenio Iglesias,
  • Ian P. Johnson,
  • Jonathan Williams-Ramirez,
  • Dina Zemlyanker,
  • Lin Tian,
  • Karthik Gopinath,
  • Mark Olchanyi,
  • Ava D. Farnan,
  • Amelia Demopoulos,
  • Matthew S. Rosen,
  • Kevin N. Sheth,
  • Adam de Havenon,
  • W. Taylor Kimberly,
  • Annabel Sorby-Adams

摘要

Portable low-field MRI offers an affordable and mobile alternative to conventional high-field scanners, enabling imaging in point-of-care and resource-limited settings. However, its lower signal-to-noise ratio, reduced resolution, and acquisition artifacts raise concerns about the accuracy of standard image registration methods. Reliable nonlinear registration is critical for a wide range of emerging applications, including frequent brain monitoring, assessment of neurodegenerative disease progression, and evaluation of treatment effects such as those associated with Alzheimer’s therapeutics. In this work, we systematically evaluated state-of-the-art registration approaches on simulated low-field scans (obtained by downsampling high-field images) and on real low-field brain MRI data. We compared three approaches that are representative of current registration methods: classical optimization (NiftyReg), learning-based registration (SynthMorph), and synthesis-based registration (SynthSR+NiftyReg). Using downsampled high-field scans, all methods performed well, achieving high Dice scores and smooth deformation fields, indicating that reduced resolution alone does not hinder registration. In contrast, real low-field data exhibited lower accuracy, primarily due to geometric distortion and other acquisition-specific artifacts. Among the tested approaches, the synthesis-based pipeline achieved the most robust performance across subjects and modalities. Overall, existing algorithms can accommodate resolution limitations, however, future methods could further enhance coregistration by explicitly addressing the distortions present in low-field MRI scans.