<p>We demonstrate the feasibility of machine-learning aided UV absorbance spectroscopy for in-process microbial contamination detection during cell therapy product (CTP) manufacturing. This method leverages a one-class support vector machine to analyse the absorbance spectra of cell cultures and predict if a sample is sterile or contaminated. This label-free technique provides a rapid output (&lt; 30&#xa0;minutes) with minimal sample preparation and volume (&lt; 1&#xa0;mL). Spiking of 7 microbial organisms into mesenchymal stromal cells supernatant aliquots from 6 commercial donors showed that contamination events could be detected at low inoculums of 10 CFUs with mean true positive and negative rates of 92.7% and 77.7% respectively. The true negative rate further improved to 92% after excluding samples from a single donor with anomalously high nicotinic acid. In cells spiked with 10 CFUs of <i>E. coli</i>, contamination was detected at the 21-hour timepoint, demonstrating comparable sensitivity to compendial USP &lt; 71 &gt; test (~ 24&#xa0;hours). We hypothesize that spectral differences between nicotinic acid and nicotinamide in the UV region are the underlying mechanisms for contamination detection. This approach can be deployed as a preliminary test during different CTP manufacturing stages, for real-time, continuous culture monitoring enabling early detection of microbial contamination, assuring safety of CTP.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Machine learning aided UV absorbance spectroscopy for microbial contamination in cell therapy products

  • Shruthi Pandi Chelvam,
  • Alice Jie Ying Ng,
  • Jiayi Huang,
  • Elizabeth Lee,
  • Maciej Baranski,
  • Derrick Yong,
  • Rohan B. H. Williams,
  • Stacy L. Springs,
  • Rajeev J. Ram

摘要

We demonstrate the feasibility of machine-learning aided UV absorbance spectroscopy for in-process microbial contamination detection during cell therapy product (CTP) manufacturing. This method leverages a one-class support vector machine to analyse the absorbance spectra of cell cultures and predict if a sample is sterile or contaminated. This label-free technique provides a rapid output (< 30 minutes) with minimal sample preparation and volume (< 1 mL). Spiking of 7 microbial organisms into mesenchymal stromal cells supernatant aliquots from 6 commercial donors showed that contamination events could be detected at low inoculums of 10 CFUs with mean true positive and negative rates of 92.7% and 77.7% respectively. The true negative rate further improved to 92% after excluding samples from a single donor with anomalously high nicotinic acid. In cells spiked with 10 CFUs of E. coli, contamination was detected at the 21-hour timepoint, demonstrating comparable sensitivity to compendial USP < 71 > test (~ 24 hours). We hypothesize that spectral differences between nicotinic acid and nicotinamide in the UV region are the underlying mechanisms for contamination detection. This approach can be deployed as a preliminary test during different CTP manufacturing stages, for real-time, continuous culture monitoring enabling early detection of microbial contamination, assuring safety of CTP.