Background <p>Multi-frequency bioelectrical impedance analysis (MFBIA) methods offer reliable and moderately accurate estimates of body composition in tightly controlled conditions (prandial and hydration status, recent exercise, time of day).</p> Objective <p>This study examined MFBIA reliability and validity in a real-world environment where these factors were not controlled.</p> Methods <p>Regional and total body composition estimates by MFBIA (InBody 770) were compared to dual-energy X-ray absorptiometry (DXA) in 1000 healthy adults (667 men; 333 women), including fat mass (FM), percent body fat (%BF), fat-free mass (FFM), and visceral adipose tissue (VAT). In subsets, reliability was determined from duplicate MFBIA and DXA obtained within 1 week, and total body water (TBW) was compared to single-frequency BIA (SFBIA).</p> Results <p>MFBIA demonstrated modest population-level agreement with DXA for total body FM (men, <i>r</i> = 0.93, bias −3.7 ± 2.6 kg; women, <i>r</i> = 0.96, bias, −1.9 ± 1.8 kg), %BF (men, <i>r</i> = 0.89, bias, −4.2 ± 3.0%; women, <i>r</i> = 0.92, bias, −2.8 ± 2.6%), and FFM (men, <i>r</i> = 0.95, bias, 3.4 ± 2.8 kg; women, <i>r</i> = 0.94, bias, 2.0 ± 2.2 kg). Regional correlations were highest for trunk FM (men, <i>r</i> = 0.92, CCC = 0.86; women <i>r</i> = 0.93, CCC = 0.93) and lowest for VAT (men, <i>r</i> = 0.74, CCC = 0.68; women, <i>r</i> = 0.74, CCC = 0.34). DXA and MFBIA regional and total assessments were highly reliable (DXA, ICC 0.990–0.998) and (MFBIA, ICC 0.987–0.995). TBW by MFBIA and SFBIA showed moderate agreement (men, <i>r</i> = 0.73, bias, −1.89 ± 3.31; women, <i>r</i> = 0.82, bias, −1.74 ± 2.01).</p> Conclusion <p>This MFBIA system was shown to have high retest reliability and, when compared to laboratory methods, provides a moderately accurate method for measuring TBW and body composition (except for VAT) in uncontrolled conditions.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Real-world assessment of Multi-Frequency Bioelectrical Impedance Analysis (MFBIA) for measuring body composition in healthy physically active populations

  • Adam W. Potter,
  • Leigh C. Ward,
  • Christopher L. Chapman,
  • William J. Tharion,
  • David P. Looney,
  • Karl E. Friedl

摘要

Background

Multi-frequency bioelectrical impedance analysis (MFBIA) methods offer reliable and moderately accurate estimates of body composition in tightly controlled conditions (prandial and hydration status, recent exercise, time of day).

Objective

This study examined MFBIA reliability and validity in a real-world environment where these factors were not controlled.

Methods

Regional and total body composition estimates by MFBIA (InBody 770) were compared to dual-energy X-ray absorptiometry (DXA) in 1000 healthy adults (667 men; 333 women), including fat mass (FM), percent body fat (%BF), fat-free mass (FFM), and visceral adipose tissue (VAT). In subsets, reliability was determined from duplicate MFBIA and DXA obtained within 1 week, and total body water (TBW) was compared to single-frequency BIA (SFBIA).

Results

MFBIA demonstrated modest population-level agreement with DXA for total body FM (men, r = 0.93, bias −3.7 ± 2.6 kg; women, r = 0.96, bias, −1.9 ± 1.8 kg), %BF (men, r = 0.89, bias, −4.2 ± 3.0%; women, r = 0.92, bias, −2.8 ± 2.6%), and FFM (men, r = 0.95, bias, 3.4 ± 2.8 kg; women, r = 0.94, bias, 2.0 ± 2.2 kg). Regional correlations were highest for trunk FM (men, r = 0.92, CCC = 0.86; women r = 0.93, CCC = 0.93) and lowest for VAT (men, r = 0.74, CCC = 0.68; women, r = 0.74, CCC = 0.34). DXA and MFBIA regional and total assessments were highly reliable (DXA, ICC 0.990–0.998) and (MFBIA, ICC 0.987–0.995). TBW by MFBIA and SFBIA showed moderate agreement (men, r = 0.73, bias, −1.89 ± 3.31; women, r = 0.82, bias, −1.74 ± 2.01).

Conclusion

This MFBIA system was shown to have high retest reliability and, when compared to laboratory methods, provides a moderately accurate method for measuring TBW and body composition (except for VAT) in uncontrolled conditions.