Background <p>Accurate assessment of spinopelvic alignment is essential for managing adult spinal deformities. This study validates the Brainlab Elements Spine Planning software for automated measurement of spinopelvic parameters, comparing it with manual methods.</p> Methods <p>Spinopelvic parameters were measured manually and with the software in 21 patients with degenerative spinal disease, including instrumented and non-instrumented spines. Accuracy, intraobserver, and interobserver reliability were evaluated using Bland–Altman plots and intraclass correlation coefficients (ICCs). Measurement times were also compared.</p> Results <p>The software showed high reliability (ICC = 1), while manual measurements ranged from fair to excellent reliability (ICC 0.44–0.99). Bland–Altman plots indicated strong agreement between automated and manual measurements, though variability was noted for certain parameters. Automated measurements were significantly faster, averaging 62 s versus 227 s in manual measurements (p &lt; 0.001), though 76.2% of cases had at least one parameter that could not be measured automatically, most frequently the sagittal vertical axis (SVA) and several coronal parameters.</p> Conclusion <p>The Brainlab Elements software provides accurate, reproducible, and time-efficient spinopelvic measurements for parameters it successfully captures. However, frequent failures in assessing SVA and coronal plane parameters automatically suggest that further refinement of the software is necessary.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Automated measurement of spinopelvic alignment parameters using a spine planning software: a validation study

  • Ahmad Almahozi,
  • Anton Früh,
  • Tarik Alp Sargut,
  • Tizian Rosenstock,
  • Claudius Jelgersma,
  • Anna L. Roethe,
  • Dimitri Tkatschenko,
  • Peter Truckenmueller,
  • Joan Alsolivany,
  • Kiarash Ferdowssian,
  • Nils Hecht,
  • Peter Vajkoczy,
  • Lars Wessels

摘要

Background

Accurate assessment of spinopelvic alignment is essential for managing adult spinal deformities. This study validates the Brainlab Elements Spine Planning software for automated measurement of spinopelvic parameters, comparing it with manual methods.

Methods

Spinopelvic parameters were measured manually and with the software in 21 patients with degenerative spinal disease, including instrumented and non-instrumented spines. Accuracy, intraobserver, and interobserver reliability were evaluated using Bland–Altman plots and intraclass correlation coefficients (ICCs). Measurement times were also compared.

Results

The software showed high reliability (ICC = 1), while manual measurements ranged from fair to excellent reliability (ICC 0.44–0.99). Bland–Altman plots indicated strong agreement between automated and manual measurements, though variability was noted for certain parameters. Automated measurements were significantly faster, averaging 62 s versus 227 s in manual measurements (p < 0.001), though 76.2% of cases had at least one parameter that could not be measured automatically, most frequently the sagittal vertical axis (SVA) and several coronal parameters.

Conclusion

The Brainlab Elements software provides accurate, reproducible, and time-efficient spinopelvic measurements for parameters it successfully captures. However, frequent failures in assessing SVA and coronal plane parameters automatically suggest that further refinement of the software is necessary.