<p>The photoacoustic spectroscopy apparatus based on an external cavity quantum cascade laser was employed to simultaneously detect four SF<sub>6</sub> decomposition by-products, i.e., SO<sub>2</sub>F<sub>2</sub>, SOF<sub>2</sub>, CF<sub>4</sub>, and SO<sub>2</sub> for the application in diagnostics of the SF<sub>6</sub> insulation equipment widely used in high-voltage transmission lines. The influence of background gases SF<sub>6</sub> and N<sub>2</sub> with the He addition in different ratios on the SOF<sub>2</sub> photoacoustic signal was investigated, with the pure SF<sub>6</sub> showing the most pronounced enhancement effect on the photoacoustic signal of the target analyte. The limit of detection of SOF<sub>2</sub> in pure SF<sub>6</sub> was estimated by comparing the limits of detection of SOF<sub>2</sub>, CF<sub>4</sub>, and SO<sub>2</sub> in the 95% SF<sub>6</sub>/5% He combination and pure SF<sub>6</sub>. The estimated detection limits for SO<sub>2</sub>F<sub>2</sub>, SOF<sub>2</sub>, CF<sub>4</sub>, and SO<sub>2</sub> were 635.4 ppb, 24.0 ppb, 10.3 ppb, and 97.8 ppb, respectively. Finally, the concentration of each decomposition component in the gas mixture of SO<sub>2</sub>F<sub>2</sub>, SOF<sub>2</sub>, CF<sub>4</sub>, and SO<sub>2</sub> in the 95% SF<sub>6</sub>/5% He background gas was quantified by multiple linear regression.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Highly Sensitive Photoacoustic Multi-Gas Detection of SF6 Decomposition Components

  • Qianhe Wei,
  • Bincheng Li,
  • Binxing Zhao,
  • Yafei Wang

摘要

The photoacoustic spectroscopy apparatus based on an external cavity quantum cascade laser was employed to simultaneously detect four SF6 decomposition by-products, i.e., SO2F2, SOF2, CF4, and SO2 for the application in diagnostics of the SF6 insulation equipment widely used in high-voltage transmission lines. The influence of background gases SF6 and N2 with the He addition in different ratios on the SOF2 photoacoustic signal was investigated, with the pure SF6 showing the most pronounced enhancement effect on the photoacoustic signal of the target analyte. The limit of detection of SOF2 in pure SF6 was estimated by comparing the limits of detection of SOF2, CF4, and SO2 in the 95% SF6/5% He combination and pure SF6. The estimated detection limits for SO2F2, SOF2, CF4, and SO2 were 635.4 ppb, 24.0 ppb, 10.3 ppb, and 97.8 ppb, respectively. Finally, the concentration of each decomposition component in the gas mixture of SO2F2, SOF2, CF4, and SO2 in the 95% SF6/5% He background gas was quantified by multiple linear regression.