Comparative analysis of AI-generated and deformed image registration contours on daily CBCT in prostate cancer radiation therapy: accuracy and dosimetric implications using commercial tools
摘要
Deep learning (DL)-based auto-segmentation has rapidly become the state-of-the-art in radiotherapy planning, significantly reducing contouring time while achieving geometric accuracy comparable to expert-derived contours [
Twenty prostate radiotherapy patients were retrospectively selected, treated with either 42.7 Gy in 7 fractions or 60 Gy in 20 fractions, and imaged on Halcyon linear accelerators using Hypersight CBCT (
AI-generated contours achieved acceptable clinical accuracy in >80% of cases, with fewer severe or medium errors compared to DIR-derived contours, which required minimal changes of 49%. Quantitative analysis demonstrated broadly comparable Dice Similarity Coefficients (DSC), Hausdorff Distance (HD), and mean surface distance (MSD) across prostate, bladder, and rectum. Organ variation on CBCT revealed larger mean centre of mass shifts and volume differences for AI, particularly in bladder contours, whereas DIR showed smaller systematic deviations. Dosimetric comparisons highlighted that prostate dose metrics were significantly different between methods, while bladder differences were mostly non-significant except at high-dose volumes, and rectum analysis revealed consistent statistically significant variations. Overall, although both methods captured daily anatomical changes, suggesting complementary strengths depending on adaptive radiotherapy application.
Conclusion:AI-generated contours for prostate radiotherapy on CBCT images demonstrate high geometric accuracy and clinical usability, requiring minimal expert correction, while DIR contours, although generally usable, show greater variability, particularly for organs subject to large anatomical changes such as the bladder and rectum. Despite similar geometric comparisons, statistically significant dosimetric differences highlight the importance of careful expert verification, especially for sensitive structures like the rectum. These findings support the integration of AI-based contouring into adaptive radiotherapy workflows to streamline clinical processes, reduce workload, and maintain treatment accuracy, while emphasizing that automated contours, whether AI- or DIR-derived, should always undergo expert review to ensure safe and effective patient care.