Purpose <p>To evaluate the impact of an AI-based, noise reduction technique for compensation of image degradation on pediatric and neonatal chest and abdomen radiography using a visual grading analysis.</p> Materials and methods <p>Forty-six consecutive cases of pediatric and neonatal chest X-rays were identified for the quality evaluation. The images underwent AI-based noise reduction processing (Intelligent NR, Canon Inc.). All the images were randomized, and were evaluated from 1 to 4 for image quality by three board-certified radiologists in consensus. A score of “1” indicated the desired anatomy or features were not seen, “2” indicated quality between one and three, “3” indicated adequate quality, and “4” indicated higher than required image quality. A Wilcoxon signed rank test was used to assess the significant difference between images from conventional noise reduction versus those from the AI-based noise reduction.</p> Results <p>The images processed with the INR(Intelligent NR) noise reduction had a higher image quality than the conventionally processed images, with a significant difference between the two groups (<i>p</i> &lt; 0.05).</p> Conclusion <p>The AI-based noise reduction technique improved the image quality of pediatric and neonatal chest and abdominal radiography significantly.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Improving image quality on pediatric and neonatal radiography using AI-based compensation for image degradation

  • So Ode,
  • Atsuko Fujikawa,
  • Atsushi Hiroishi,
  • Yuki Saito,
  • Takao Tanuma,
  • Daigo Suzuki,
  • Yuichi Sasaki,
  • Hidefumi Mimura

摘要

Purpose

To evaluate the impact of an AI-based, noise reduction technique for compensation of image degradation on pediatric and neonatal chest and abdomen radiography using a visual grading analysis.

Materials and methods

Forty-six consecutive cases of pediatric and neonatal chest X-rays were identified for the quality evaluation. The images underwent AI-based noise reduction processing (Intelligent NR, Canon Inc.). All the images were randomized, and were evaluated from 1 to 4 for image quality by three board-certified radiologists in consensus. A score of “1” indicated the desired anatomy or features were not seen, “2” indicated quality between one and three, “3” indicated adequate quality, and “4” indicated higher than required image quality. A Wilcoxon signed rank test was used to assess the significant difference between images from conventional noise reduction versus those from the AI-based noise reduction.

Results

The images processed with the INR(Intelligent NR) noise reduction had a higher image quality than the conventionally processed images, with a significant difference between the two groups (p < 0.05).

Conclusion

The AI-based noise reduction technique improved the image quality of pediatric and neonatal chest and abdominal radiography significantly.