<p><i>Streptococcus pneumoniae</i> can be responsible for severe infections, especially in patients with primary antibody deficiencies like selective anti-polysaccharide antibodies deficiency (SPAD). The reference method recommaned by the World Health Organization for assessment of anti-pneumococcal capsular polysaccharides (PCPs) IgG antibodies is a standardized serotype-specific ELISA (WHO-SSA), but this manual method is time-consuming and limit the number of evaluated PCPs. We aim to evaluate the performance values of a multiplex assay based on electrochemiluminescence (ECL-plex). A panel of 164 sera from 82 patients sampled before and 4–8 weeks after immunization by the 23-valent pneumococcal polysaccharide vaccine (PPV23) were assessed by the reference WHO-SSA (7 to 13 serotypes) and by an 18-plex ECL assay (18 serotypes). All patients had normal serum Ig/subclasses levels and were classified as good (<i>n</i> = 43) or poor responders (<i>n</i> = 39, i.e. SPAD patients) according to the American Academy of Asthma, Allergy and Immunology’s (AAAAI) current guidelines. We observed excellent correlations between the two methods for anti-PCPs titers against 7 serotypes (<i>r</i> = 0.88 [95% CI: 0.87–0.90], <i>n</i> = 124 sera) and 13 serotypes (<i>r</i> = 0.87 [0.87–0.89], <i>n</i> = 40 sera). Using the AAAAI’s guidelines for interpretation, the test performance of the 18-plex ECL assay for SPAD diagnosis showed a sensitivity of 95% and specificity of 84%, positive and negative predictive values of 84% and 95%, respectively. The percentage of agreement was 89% between the SSA and the 18-plex ECL assay. The 18-plex ECL assay is a reliable, rapid, and simple method for evaluating anti-PCPs response and screening for SPAD diagnosis.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Evaluation of a Multiplex Electrochemiluminescence Assay for Detection of Anti-Pneumococcal Antibodies in the Diagnosis of Selective Polysaccharide Antibody Deficiency

  • Nicolas Perrard,
  • Aurore Collet,
  • Sarah Stabler,
  • Sandrine Poizot,
  • Myriam Labalette,
  • Gatien Durand,
  • Frédéric Batteux,
  • Floriane Mirgot,
  • Benjamin Lopez,
  • Sylvain Dubucquoi,
  • Lucie Chevrier,
  • Guillaume Lefèvre

摘要

Streptococcus pneumoniae can be responsible for severe infections, especially in patients with primary antibody deficiencies like selective anti-polysaccharide antibodies deficiency (SPAD). The reference method recommaned by the World Health Organization for assessment of anti-pneumococcal capsular polysaccharides (PCPs) IgG antibodies is a standardized serotype-specific ELISA (WHO-SSA), but this manual method is time-consuming and limit the number of evaluated PCPs. We aim to evaluate the performance values of a multiplex assay based on electrochemiluminescence (ECL-plex). A panel of 164 sera from 82 patients sampled before and 4–8 weeks after immunization by the 23-valent pneumococcal polysaccharide vaccine (PPV23) were assessed by the reference WHO-SSA (7 to 13 serotypes) and by an 18-plex ECL assay (18 serotypes). All patients had normal serum Ig/subclasses levels and were classified as good (n = 43) or poor responders (n = 39, i.e. SPAD patients) according to the American Academy of Asthma, Allergy and Immunology’s (AAAAI) current guidelines. We observed excellent correlations between the two methods for anti-PCPs titers against 7 serotypes (r = 0.88 [95% CI: 0.87–0.90], n = 124 sera) and 13 serotypes (r = 0.87 [0.87–0.89], n = 40 sera). Using the AAAAI’s guidelines for interpretation, the test performance of the 18-plex ECL assay for SPAD diagnosis showed a sensitivity of 95% and specificity of 84%, positive and negative predictive values of 84% and 95%, respectively. The percentage of agreement was 89% between the SSA and the 18-plex ECL assay. The 18-plex ECL assay is a reliable, rapid, and simple method for evaluating anti-PCPs response and screening for SPAD diagnosis.