Purpose <p>Intense pulsed light (IPL) therapy is an approach to ameliorate meibomian gland dysfunction (MGD). A novel IPL system for the treatment of MGD is the LacryStim device. The aim of this study is to investigate the safety and efficacy of the LacryStim device in patients with MGD.</p> Methods <p>This study was conducted in a prospective, randomized, intra-individual sham-controlled, double-masked design. Three treatment visits and three study visits at baseline and after three and six months were performed. The randomized study eye (SE) was treated using 8&#xa0;J/cm<sup>2</sup>. The contralateral control eye (CE) was sham-treated with 1&#xa0;J/cm<sup>2</sup>. The primary outcome was the mean difference of meibum quality and expressibility between SEs and CEs at the six months visit.</p> Results <p>Fifty eyes of 25 patients were included in the study. While no significant differences between both groups regarding meibum quality and expressibility could be detected at the six months visit, there was a tendency towards better expressibility in both groups after six months (SEs: − 0.4, <i>p</i> = 0.07; CEs: − 0.4, <i>p</i> = 0.04). Additionally, a significant improvement in numeric rating scale in both eyes and OSDI score (both <i>p</i> &lt; 0.05) at all follow-up visits and no device-related adverse events were observed.</p> Conclusion <p>To our knowledge this is the first prospective evaluation of a novel IPL device. There was a significant amelioration of subjective complaints and a tendency towards better expressibility in both study and control eyes. An intra-individual design might compound the individual analysis of either eyes in studies investigating IPL.</p> <p>ClinicalTrials.gov Identifier: NCT05089591.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Safety and efficacy of a novel intense pulsed light system in patients with meibomian gland dysfunction: a randomized, double-masked, intra-individual controlled study

  • Andreas Schlatter,
  • Stefan Palkovits,
  • Sahand Amir-Asgari,
  • Manuel Ruiss,
  • Johannes Zeilinger,
  • Andreas Rotter,
  • Doreen Schmidl,
  • Oliver Findl

摘要

Purpose

Intense pulsed light (IPL) therapy is an approach to ameliorate meibomian gland dysfunction (MGD). A novel IPL system for the treatment of MGD is the LacryStim device. The aim of this study is to investigate the safety and efficacy of the LacryStim device in patients with MGD.

Methods

This study was conducted in a prospective, randomized, intra-individual sham-controlled, double-masked design. Three treatment visits and three study visits at baseline and after three and six months were performed. The randomized study eye (SE) was treated using 8 J/cm2. The contralateral control eye (CE) was sham-treated with 1 J/cm2. The primary outcome was the mean difference of meibum quality and expressibility between SEs and CEs at the six months visit.

Results

Fifty eyes of 25 patients were included in the study. While no significant differences between both groups regarding meibum quality and expressibility could be detected at the six months visit, there was a tendency towards better expressibility in both groups after six months (SEs: − 0.4, p = 0.07; CEs: − 0.4, p = 0.04). Additionally, a significant improvement in numeric rating scale in both eyes and OSDI score (both p < 0.05) at all follow-up visits and no device-related adverse events were observed.

Conclusion

To our knowledge this is the first prospective evaluation of a novel IPL device. There was a significant amelioration of subjective complaints and a tendency towards better expressibility in both study and control eyes. An intra-individual design might compound the individual analysis of either eyes in studies investigating IPL.

ClinicalTrials.gov Identifier: NCT05089591.