Purpose <p>This study aimed to establish a reference data set of light-adapted full-field electroretinograms (ffERG) of healthy primary and secondary schoolchildren and investigate the relationship among refraction, ocular biometry, and ffERG.</p> Methods <p>Healthy children aged between 6 and 17&#xa0;years were recruited. Cycloplegic spherical equivalent refraction (SER), axial length (AL) and keratometry (K) measurements were performed. Standardized ffERGs, including light-adapted flash and 30-Hz flicker, were measured using a portable device with skin electrodes. The percentiles of peak time and amplitude of a- and b-waves and 30-Hz flicker of ffERG waveform were reported, and their relationships with age, SER, AL, K were investigated.</p> Results <p>Among the 445 participants (45.4% female), the SER (mean ± standard deviation) was − 0.72 ± 2.06 D and AL was 23.56 ± 1.15&#xa0;mm. The 90% confidence interval (CI) of 5th reference limit of amplitudes of a- and b-wave were 5.0–5.8&#xa0;µV and 15.0–17.6&#xa0;µV, while 95th reference limit of peak times were 13.2–13.4&#xa0;ms and 29.8–30.0&#xa0;ms, respectively. The amplitude and peak time of the 30-Hz flicker waveform were 21.5–23.9&#xa0;µV and 26.0–26.2&#xa0;ms, respectively. In general, more myopic SER, and longer AL were associated with delayed and reduced ffERG waveforms. Older age and male sex were weakly correlated with ffERGs with minimal clinical significance.</p> Conclusions <p>A reference data set of light-adapted ffERG in children and adolescents was established for clinical and research purposes.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Reference ranges of light-adapted full-field electroretinogram and associated factors in a large cohort of healthy school-aged children and adolescents

  • Sonia Seen-hang Chan,
  • Kai Yip Choi,
  • Natalie Yu-yan Chan,
  • Vivian Wai Ying Lo,
  • Angela Hing-yiu Hung,
  • Henry Ho-lung Chan

摘要

Purpose

This study aimed to establish a reference data set of light-adapted full-field electroretinograms (ffERG) of healthy primary and secondary schoolchildren and investigate the relationship among refraction, ocular biometry, and ffERG.

Methods

Healthy children aged between 6 and 17 years were recruited. Cycloplegic spherical equivalent refraction (SER), axial length (AL) and keratometry (K) measurements were performed. Standardized ffERGs, including light-adapted flash and 30-Hz flicker, were measured using a portable device with skin electrodes. The percentiles of peak time and amplitude of a- and b-waves and 30-Hz flicker of ffERG waveform were reported, and their relationships with age, SER, AL, K were investigated.

Results

Among the 445 participants (45.4% female), the SER (mean ± standard deviation) was − 0.72 ± 2.06 D and AL was 23.56 ± 1.15 mm. The 90% confidence interval (CI) of 5th reference limit of amplitudes of a- and b-wave were 5.0–5.8 µV and 15.0–17.6 µV, while 95th reference limit of peak times were 13.2–13.4 ms and 29.8–30.0 ms, respectively. The amplitude and peak time of the 30-Hz flicker waveform were 21.5–23.9 µV and 26.0–26.2 ms, respectively. In general, more myopic SER, and longer AL were associated with delayed and reduced ffERG waveforms. Older age and male sex were weakly correlated with ffERGs with minimal clinical significance.

Conclusions

A reference data set of light-adapted ffERG in children and adolescents was established for clinical and research purposes.