Purpose <p>Visual acuity (VA) assessment is crucial for evaluating visual function, but inter-clinic VA measurement variations were observed in patients with epiretinal membrane (ERM). Inter-clinic variations in acuity chart luminance exist because the International Organization for Standardization (ISO) recommends a range of 80–320 cd/m<sup>2</sup> for VA charts. We investigated the effects of luminance within this ISO range on the VA in patients with ERM.</p> Study design <p>Prospective observational study.</p> Methods <p>We analyzed data from 187 patients with ERM and a normal fellow eye with a decimal VA of ≥1.0 at 320 cd/m<sup>2</sup> OU. Baseline VA was measured at 320 cd/m<sup>2</sup>, then reduced in 0.1-log unit steps using neutral density filters. Metamorphopsia was evaluated with the Amsler Grid Test. The luminance causing a two-line VA reduction from 320 cd/m<sup>2</sup> was measured before and after ERM removal.</p> Results <p>In normal fellow eyes, a two-line VA reduction occurred at approximately 40 cd/m<sup>2</sup>, indicating stable VA within the normal ISO range, while in affected eyes, it happened at around 150 cd/m<sup>2</sup>. The luminance difference causing a two-line VA reduction between eyes with and without metamorphopsia was not significant. For surgical patients, the luminance causing a two-line VA reduction decreased over time: from 153 cd/m<sup>2</sup> preoperatively to 36 cd/m<sup>2</sup> at 6 months postoperatively.</p> Conclusions <p>Considering the luminance is critical when measuring VA in patients with ERM.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

The effect of luminance on visual acuity measurement in eyes with epiretinal membrane

  • Yui Nariai,
  • Masayuki Horiguchi,
  • Tadashi Mizuguchi,
  • Ryota Sakurai,
  • Yasuki Ito

摘要

Purpose

Visual acuity (VA) assessment is crucial for evaluating visual function, but inter-clinic VA measurement variations were observed in patients with epiretinal membrane (ERM). Inter-clinic variations in acuity chart luminance exist because the International Organization for Standardization (ISO) recommends a range of 80–320 cd/m2 for VA charts. We investigated the effects of luminance within this ISO range on the VA in patients with ERM.

Study design

Prospective observational study.

Methods

We analyzed data from 187 patients with ERM and a normal fellow eye with a decimal VA of ≥1.0 at 320 cd/m2 OU. Baseline VA was measured at 320 cd/m2, then reduced in 0.1-log unit steps using neutral density filters. Metamorphopsia was evaluated with the Amsler Grid Test. The luminance causing a two-line VA reduction from 320 cd/m2 was measured before and after ERM removal.

Results

In normal fellow eyes, a two-line VA reduction occurred at approximately 40 cd/m2, indicating stable VA within the normal ISO range, while in affected eyes, it happened at around 150 cd/m2. The luminance difference causing a two-line VA reduction between eyes with and without metamorphopsia was not significant. For surgical patients, the luminance causing a two-line VA reduction decreased over time: from 153 cd/m2 preoperatively to 36 cd/m2 at 6 months postoperatively.

Conclusions

Considering the luminance is critical when measuring VA in patients with ERM.