Purpose <p>To investigate the safety and aesthetic outcomes of silicone buckle #506 for anophthalmic socket reconstruction.</p> Study Design <p>Retrospective cohort study.</p> Methods <p>Patients who underwent enucleation or evisceration at the Osaka University Hospital were retrospectively analyzed. Patients under 18 years old, with &lt; 3 months follow-up, without a prosthesis, or with bilateral surgery were excluded. Aesthetic outcomes were assessed using standardized photographs taken 3 months postoperatively, scored by five independent ophthalmologists for upper eyelid sulcus deepening asymmetry. Scoring was categorized as 0: no noticeable side difference, 1: slight asymmetry, and 2: marked asymmetry characterized by upper eyelid sulcus deepening.</p> Results <p>Fifty-nine patients (60 sockets) were analyzed. Thirty-three received silicone buckle implants (1–4 buckles). Implant exposure occurred in two patients (6.1%). Aesthetic scores were assessed in 48 patients. Mean aesthetic scores were 0.97 (no implant), 0.78 (one buckle), 0.68 (two buckles), and 0.42 (three/four buckles) (<i>p</i> = 0.123).</p> Conclusions <p>Silicone buckle #506 appears to be a safe and feasible option for anophthalmic socket reconstruction. Further studies are needed to optimize aesthetic outcomes and determine the ideal number of buckles.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Safety and aesthetic outcomes of silicone buckles as an implant for anophthalmic sockets: a decade of experience in Japan

  • Rikako Iwasaki,
  • Yoshiyuki Kitaguchi,
  • Takeshi Morimoto,
  • Hiroshi Shimojo,
  • Takahiro Fujino,
  • Shimpei Komoto,
  • Kohji Nishida

摘要

Purpose

To investigate the safety and aesthetic outcomes of silicone buckle #506 for anophthalmic socket reconstruction.

Study Design

Retrospective cohort study.

Methods

Patients who underwent enucleation or evisceration at the Osaka University Hospital were retrospectively analyzed. Patients under 18 years old, with < 3 months follow-up, without a prosthesis, or with bilateral surgery were excluded. Aesthetic outcomes were assessed using standardized photographs taken 3 months postoperatively, scored by five independent ophthalmologists for upper eyelid sulcus deepening asymmetry. Scoring was categorized as 0: no noticeable side difference, 1: slight asymmetry, and 2: marked asymmetry characterized by upper eyelid sulcus deepening.

Results

Fifty-nine patients (60 sockets) were analyzed. Thirty-three received silicone buckle implants (1–4 buckles). Implant exposure occurred in two patients (6.1%). Aesthetic scores were assessed in 48 patients. Mean aesthetic scores were 0.97 (no implant), 0.78 (one buckle), 0.68 (two buckles), and 0.42 (three/four buckles) (p = 0.123).

Conclusions

Silicone buckle #506 appears to be a safe and feasible option for anophthalmic socket reconstruction. Further studies are needed to optimize aesthetic outcomes and determine the ideal number of buckles.