Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 1): apparent diffusion coefficient standardization in a diffusion MRI phantom
摘要
To validate multi-site and multi-vendor ADC measurements using the QIBA/NIST diffusion MRI phantom at room temperature.
Materials and methodsADC measurements were performed on 12 scanners (evenly split between 1.5 and 3 T) from three vendors at five sites and compared with reference values at room temperature. We adopted Pearson’s correlation (r) and accuracy error for comparison with reference values; within scanner coefficient of variation (CVintra%) for intra-session repeatability and inter-scanner for agreement (CVinter%); Bland–Altman plots and precision error for short-term reproducibility; generalized linear mixed models and post-hoc tests (
Temperature adjusted ADCs were well correlated with NIST reference values (r
Despite using abdominal receive coils and room temperature measurements, all quantitative parameters were within literature findings. High accuracy, repeatability and precision within the renal physiologic range support the feasibility of scanner evaluation using QIBA standardization process for diffusion measurements in renal studies.