Change in composition of blood after passing through circuit of automated device versus manual clinical method for neonatal exchange transfusions
摘要
We previously developed and published a novel device to automate double volume exchange transfusions (DVET) via the umbilical venous catheter (UVC) in neonates. Prior to initiating clinical trials, we aimed to demonstrate that alterations in blood composition after passage through our device’s circuit were not greater than those seen with the conventional manual method. We performed five ex vivo simulations of neonatal DVET using donor blood in standard blood bank bags and our automated device. Samples were obtained from the blood bag at baseline and from the distal end of the circuit after completion of 50% and 100% of the simulated DVET. These were analyzed for hematological, biochemical, and blood gas parameters, osmolality, and culture. For comparison, four manual DVETs were similarly performed, with corresponding sample collection and analyses. Deviations of laboratory parameters at the 50% and 100% time points that were > ± 25% from baseline were classified as major. The proportion of major deviations was compared between automated and manual DVETs. Automated DVETs demonstrated significantly fewer major deviations than manual DVETs in hematological parameters [15/156 (9.61%) vs. 28/124 (22.58%), p = 0.0048], biochemical parameters [0/120 (0%) vs. 32/88 (36.36%), p < 0.0001], and blood gas values [11/50 (22.0%) vs. 20/40 (50.0%), p = 0.011]. Notably, the blood used in automated DVETs was significantly older than that used in the manual group.
Conclusion: Automated DVET using our novel device results in fewer compositional deviations from baseline compared to conventional manual DVET, supporting its potential clinical utility.