Aims <p>The number of patients with cardiac implantable electronic devices (CIEDs) is increasing. However, there is limited experience regarding vasodilator-stress cardiovascular magnetic resonance (CMR) and resulting device artifacts on perfusion images. The aim of this study was to determine CMR image quality in patients with different CIED types for CMR-based perfusion stress testing.</p> Methods and results <p>A total of 156 patients with active CIEDs underwent CMR on a 1.5-Tesla scanner. Both conventional steady-state-free-precession (SSFP) and modified spoiled gradient-echo (sGE) protocols under stress and resting conditions were used to evaluate image artifacts in a 16-segment segment model of the heart. The study group comprised 39% conventional pacemaker (PM), 4% cardiac resynchronization therapy-pacemaker (CRT-P), 38% conventional implantable cardioverter-defibrillator (ICD), 6% cardiac resynchronization therapy-ICD (CRT-D), and 13% subcutaneous ICD (S-ICD) patients. PM-carriers showed only minor image artifacts in both perfusion protocols. Artifacts caused by ICDs were predominantly located in the left-ventricular (LV) inferolateral and anterior segments. S-ICDs showed the highest extent of artifacts with an anterolateral accentuation. The artifact extent was significantly reduced when sGE-based perfusion was used compared to SSFP-based sequences. 69% of the patients received a stress-perfusion protocol, and elective coronary angiography confirmed the presence of coronary stenosis in three cases. No major safety-relevant issues occurred.</p> Conclusion <p>Myocardial perfusion imaging by CMR is safe and feasible with moderate-to-high image quality in patients with all types of CIEDs, including non-conditional devices, ICDs, and S-ICDs. A sGE-based perfusion protocol should be preferred in patients with left-sided ICDs, CRT-Ds, or S-ICDs.</p> Graphical abstract <p></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

First-pass perfusion imaging using cardiovascular magnetic resonance in patients with various cardiac implantable electronic devices

  • Claudia Meier,
  • Michael Bietenbeck,
  • Maria Theofanidou,
  • Volker Vehof,
  • Philipp Stalling,
  • Dennis Korthals,
  • Bishwas Chamling,
  • Misael Estepa,
  • Patrick Doeblin,
  • Sebastian Kelle,
  • Ali Yilmaz

摘要

Aims

The number of patients with cardiac implantable electronic devices (CIEDs) is increasing. However, there is limited experience regarding vasodilator-stress cardiovascular magnetic resonance (CMR) and resulting device artifacts on perfusion images. The aim of this study was to determine CMR image quality in patients with different CIED types for CMR-based perfusion stress testing.

Methods and results

A total of 156 patients with active CIEDs underwent CMR on a 1.5-Tesla scanner. Both conventional steady-state-free-precession (SSFP) and modified spoiled gradient-echo (sGE) protocols under stress and resting conditions were used to evaluate image artifacts in a 16-segment segment model of the heart. The study group comprised 39% conventional pacemaker (PM), 4% cardiac resynchronization therapy-pacemaker (CRT-P), 38% conventional implantable cardioverter-defibrillator (ICD), 6% cardiac resynchronization therapy-ICD (CRT-D), and 13% subcutaneous ICD (S-ICD) patients. PM-carriers showed only minor image artifacts in both perfusion protocols. Artifacts caused by ICDs were predominantly located in the left-ventricular (LV) inferolateral and anterior segments. S-ICDs showed the highest extent of artifacts with an anterolateral accentuation. The artifact extent was significantly reduced when sGE-based perfusion was used compared to SSFP-based sequences. 69% of the patients received a stress-perfusion protocol, and elective coronary angiography confirmed the presence of coronary stenosis in three cases. No major safety-relevant issues occurred.

Conclusion

Myocardial perfusion imaging by CMR is safe and feasible with moderate-to-high image quality in patients with all types of CIEDs, including non-conditional devices, ICDs, and S-ICDs. A sGE-based perfusion protocol should be preferred in patients with left-sided ICDs, CRT-Ds, or S-ICDs.