Breast density assessment via quantitative sound-speed measurement using conventional ultrasound transducers
摘要
The aim is to assess the feasibility and accuracy of a novel quantitative ultrasound (US) method based on global speed-of-sound (g-SoS) measurement using conventional US machines, for breast density assessment in comparison to mammographic ACR (m-ACR) categories.
Materials and methodsIn a prospective study, g-SoS was assessed in the upper-outer breast quadrant of 100 women, with 92 of them also having m-ACR assessed by two radiologists across the entire breast. For g-SoS, ultrasonic waves were transmitted from varying transducer locations and the image misalignments between these were then related analytically to breast SoS. To test reproducibility, two consecutive g-SoS acquisitions each were taken at two similar breast locations by the same operator.
ResultsMeasurements were found highly repeatable, with a mean absolute difference ± standard deviation of 3.16 ± 3.79 m/s. Multiple measurements were combined yielding a single g-SoS estimate per each patient, which strongly correlated to m-ACR categories (Spearman’s = 0.773). The g-SoS values for categories A-D were 1459.6 ± 0.74, 1475.6 ± 15.92, 1515.6 ± 27.10, and 1545.7 ± 20.62, with all groups (except A-B) being significantly different from each other. Dense breasts (m-ACR C&D) were classified with 100% specificity at 78% sensitivity, with an area under the curve (AUC) of 0.931. Extremely dense breasts (m-ACR D) were classified with 100% sensitivity at 77.5% specificity (AUC = 0.906).
ConclusionQuantitative g-SoS measurement of the breast was shown feasible and repeatable using conventional US machines, with values correlating strongly with m-ACR assessments.