Purpose <p>To evaluate the feasibility of developing an artificial intelligence application for real-time specimen adequacy assessment during ultrasound-guided fine needle aspiration of the thyroid gland using small, locally derived datasets and open-source AI tools.</p> Materials and Methods <p>1,244 photomicrographic images of thyroid needle aspiration slides were used to develop training and testing datasets for a YOLOv5 convolutional neural network model. In silico performance was tested for standard metrics. The model was deployed in an application featuring a live, augmented reality microscopy display. Hands-on testing was performed using a set of 50 blinded patient specimens tested in duplicate (N=100) by human observers with no prior formal training in pathology.</p> Results <p>On an independent static image test set, the CNN achieved an AUROC of 0.954 (95% CI, 0.922–0.986) using images acquired with a dedicated microscope camera, with sensitivity 0.969, specificity 0.933, and accuracy 95.5%. Using images acquired with a smartphone camera, performance improved to an AUROC of 0.983 (95% CI, 0.965–1.000), with sensitivity 0.980, specificity 1.000, and accuracy 98.8%. In a simulated clinical workflow, human users assisted by the augmented reality AI tool demonstrated 96% concordance with the reference standard FNA adequacy assessment.</p> Conclusion <p>An augmented reality microscopy tool based on a small-model CNN showed feasibility for performing real-time assessment of thyroid FNA specimen adequacy. Such a tool can enable telecytologic support of FNA assessment in radiologic facilities without an on-site pathologist.</p> Graphical Abstract <p></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Feasibility of an Augmented Reality Artificial Intelligence Tool for Real-Time Assessment of Thyroid Fine-Needle Aspiration Adequacy

  • Paul M. Gibbs,
  • Nguyen Dat Xuan,
  • Gladys M. Vargas,
  • Michelle J. Gibbs,
  • Pranav Sudheesh,
  • Emma K. Fitzpatrick,
  • Owen R. Tapia,
  • Josh Zhong,
  • Julie B. Steele

摘要

Purpose

To evaluate the feasibility of developing an artificial intelligence application for real-time specimen adequacy assessment during ultrasound-guided fine needle aspiration of the thyroid gland using small, locally derived datasets and open-source AI tools.

Materials and Methods

1,244 photomicrographic images of thyroid needle aspiration slides were used to develop training and testing datasets for a YOLOv5 convolutional neural network model. In silico performance was tested for standard metrics. The model was deployed in an application featuring a live, augmented reality microscopy display. Hands-on testing was performed using a set of 50 blinded patient specimens tested in duplicate (N=100) by human observers with no prior formal training in pathology.

Results

On an independent static image test set, the CNN achieved an AUROC of 0.954 (95% CI, 0.922–0.986) using images acquired with a dedicated microscope camera, with sensitivity 0.969, specificity 0.933, and accuracy 95.5%. Using images acquired with a smartphone camera, performance improved to an AUROC of 0.983 (95% CI, 0.965–1.000), with sensitivity 0.980, specificity 1.000, and accuracy 98.8%. In a simulated clinical workflow, human users assisted by the augmented reality AI tool demonstrated 96% concordance with the reference standard FNA adequacy assessment.

Conclusion

An augmented reality microscopy tool based on a small-model CNN showed feasibility for performing real-time assessment of thyroid FNA specimen adequacy. Such a tool can enable telecytologic support of FNA assessment in radiologic facilities without an on-site pathologist.