Purpose <p>Efforts to reduce the radiation burden of PET/CT have driven the increasing development of AI-based CT-less PET imaging techniques. However, comprehensive clinical evaluations of these approaches remain limited. This study aimed to rigorously assess whether deep learning (DL)-based PET reconstruction can eliminate the need for CT-derived attenuation and scatter correction while maintaining image quality sufficient for reliable clinical diagnosis.</p> Methods <p>In this dual-center retrospective analysis, raw PET/CT data from 359 patients were evaluated across 4 scanners and 4 tracers. Each dataset underwent four reconstruction approaches: (1) CT-based attenuation and scatter correction (CT-ASC, reference standard); (2) conventional 2D-DL; (3) conventional 3D-DL; and (4) our novel Decomposition-based DL algorithm. Diagnostic quality of reconstructed images was systematically assessed via visual scoring (5-point Likert scale), diagnostic accuracy (lesion-based false-positive/negative rates), and semi-quantitative metrics (SUVmax consistency).</p> Results <p>Visual analysis demonstrated the superior performance of Decomposition-based DL compared to conventional 2D-DL and 3D-DL (<i>p</i> &lt; 0.001 for all comparisons). Furthermore, the proposed method exhibited the lowest false-negative and false-positive rates (0.56% false positives with SIEMENS Vision 600; zero rates in other cases). Semi-quantitative analysis showed that although Decomposition-based DL did not consistently yield the lowest mean absolute percentage error values compared to controls, it maintained strong agreement with CT-ASC in most cases.</p> Conclusion <p>This dual-center study demonstrates that decomposition-based DL CT-free PET imaging outperforms conventional DL methods, achieving diagnostic accuracy comparable to CT-based attenuation correction in most cases. This clinical evaluation provides valuable insights to guide further methodological development and support clinical translation of CT-free PET imaging.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Clinical evaluation of deep learning-based CT-free PET reconstruction image: a dual-center study

  • Hangxing Chunyu,
  • Yizhou Chen,
  • Song Xue,
  • Xinyu Zhang,
  • Ying Miao,
  • Rui Guo,
  • Biao Li,
  • Kuangyu Shi

摘要

Purpose

Efforts to reduce the radiation burden of PET/CT have driven the increasing development of AI-based CT-less PET imaging techniques. However, comprehensive clinical evaluations of these approaches remain limited. This study aimed to rigorously assess whether deep learning (DL)-based PET reconstruction can eliminate the need for CT-derived attenuation and scatter correction while maintaining image quality sufficient for reliable clinical diagnosis.

Methods

In this dual-center retrospective analysis, raw PET/CT data from 359 patients were evaluated across 4 scanners and 4 tracers. Each dataset underwent four reconstruction approaches: (1) CT-based attenuation and scatter correction (CT-ASC, reference standard); (2) conventional 2D-DL; (3) conventional 3D-DL; and (4) our novel Decomposition-based DL algorithm. Diagnostic quality of reconstructed images was systematically assessed via visual scoring (5-point Likert scale), diagnostic accuracy (lesion-based false-positive/negative rates), and semi-quantitative metrics (SUVmax consistency).

Results

Visual analysis demonstrated the superior performance of Decomposition-based DL compared to conventional 2D-DL and 3D-DL (p < 0.001 for all comparisons). Furthermore, the proposed method exhibited the lowest false-negative and false-positive rates (0.56% false positives with SIEMENS Vision 600; zero rates in other cases). Semi-quantitative analysis showed that although Decomposition-based DL did not consistently yield the lowest mean absolute percentage error values compared to controls, it maintained strong agreement with CT-ASC in most cases.

Conclusion

This dual-center study demonstrates that decomposition-based DL CT-free PET imaging outperforms conventional DL methods, achieving diagnostic accuracy comparable to CT-based attenuation correction in most cases. This clinical evaluation provides valuable insights to guide further methodological development and support clinical translation of CT-free PET imaging.